Calibration Specifications
1.Requirement from side view
Observed from the side of the device, the ink line shall be positioned at the exact center of the recovery port to ensure stable ink recovery.

2.Requirement from top view
Observed from the top-down perspective, the ink line shall be located at the 1/4 section of the recovery port on the side close to the deflection plate.

Optional Operation Tips
If the ink line deviates from the above standard positions, you can fine-tune the corresponding adjustment knob of the nozzle to correct the ink trajectory. Recheck the position from both two perspectives after adjustment until the specifications are met, so as to prevent faults such as ink leakage, poor ink recovery and broken printing lines.